7007
|
6054
|
6055
|
Range
|
–5 °C to 110 °C |
50 °C to 300 °C |
200 °C to 550 °C |
Temperature Stability |
± 0.001 °C at 0 °C (water/ethanol)
± 0.001 °C at 25 °C (water)
± 0.003 °C at 100 °C (oil 5012) |
± 0.003 °C at 100 °C (oil 5012)
± 0.005 °C at 300 °C (oil 5017) |
± 0.003 °C at 200 °C (salt)
± 0.01 °C at 550 °C (salt) |
Temperature Uniformity |
± 0.004 °C at 0 °C (ethanol)
± 0.007 °C at 100 °C (oil 5012) |
± 0.007 °C at 100 °C (oil 5012)
± 0.015 °C at 300 °C (oil 5017) |
± 0.005 °C at 200 °C (salt)
± 0.010 °C at 550 °C (salt) |
Temperature Setting
|
Digital display with push-button data entry |
Set-Point Resolution
|
0.002 °C, high res. 0.00003 °C |
0.01 °C, high res. 0.00018 °C |
Display Temperature Resolution
|
0.01 °C |
Digital Setting Accuracy
|
± 1 °C |
Digital Setting Repeatability
|
± 0.005 °C |
± 0.01 °C |
Heaters
|
250 to 1000 W |
250 to 1000 W |
260 to 2080 W |
Working Area
|
178 mm dia. (7 in) |
196 mm dia. (7.7 in) |
107 mm dia. (4.2 in) |
Depth
|
24 in |
24 in |
24 in |
Wetted Parts
|
304 stainless steel |
Power |
230 V ac (± 10 %), 50 or 60 Hz, 15 A
(Specify frequency, contact Fluke Calibration if CE
mark required.) |
230 V ac (± 10 %), 50/60 Hz,
15 A max |
230 V ac (± 10 %), 50/60 Hz, 15 A |
Volume |
42 liters (11.2 gal) |
50 liters (13.2 gal) |
19.8 liters
(5.2 gal, 43 kg [95 lb] of bath salt) |
Size (D x W x H) |
470 x 775 x 194 mm to working surface (18.5 x 30.5 x 47 in), 1397 mm (55 in) to top of stir motor, 914 mm (36 in) to control panel |
572 x 762 x 1219 mm to working surface (22.5 x 30 x 48 in), 1422 mm (56 in) to top of stir motor box, 914 mm (36 in) to control panel |
572 x 775 x 1219 mm to working surface (22.5 x 30.5 x 48 in), 1524 mm (60 in) to top of stir motor box, 914 mm (36 in) to control panel |
Distance from Line of Sight to Top of Fluid
|
9.5 mm (3/8 in) |
15.9 mm (5/8 in) |
Automation |
Use the 1586A Super-DAQ to automate temperature sensor calibration. For IEEE-488, add the 2001-IEEE to the automation package. |
Weight
|
70.8 kg (156 lb) |